13
Long-press
(3s)
In measure mode, keep one of the two keys pressed to change the temperature
in MTC mode (manual compensation, without probe). When the value starts to
flash, the user can change the temperature value by entering the correct one and
confirming with
.
IMPORTANT:
•
When the Sleep mode is active (by default after two minutes of inactivity of the instrument) press
any key to reactivate the brightness of the display.
•
Only at this point do the keys regain their function.
•
Inputs / Outputs connections
Use original accessories guaranteed by the manufacturer only.
If necessary, contact your local distributor.
OXY 70 Vio top panel
READ THE MANUAL BEFORE PROCEEDING TO CONNECT THE PROBES OR PERIPHERALS
•
Symbols and icons on the display
Symbol
Description
Symbol
Description
M+
Number of data stored in Data Logger
mode on instrumental memory
Error in measurement or calibration
Instrument connected to the D
software
FIXED: Automatic Data Logger set
INTERMITTENT: Automatic Data
Logger in operation
6 pin connector
for optical
sensor
Micro USB port
for
connection to PC and
power supply