23
After the maintenance make a new
Calibration of the oxygen sensor.
Data Logger Function
This series of devices has the possibility of recording values in GLP format on the instrument's internal
memory.
•
The instrument can save up to 1000 data in total. Once the memory is finished, the values are NOT
overwritten. In measure mode, the number of data stored for that parameter appears next to the
M+
icon.
•
It is possible to recall and consult the values on the display or download them to a PC using the
appropriate software.
•
If you have the possibility to work directly connected to the PC, the data are automatically saved on
the software without having memory limitations.
•
Recordings can be acquired
manually
(MANUAL) or
automatically at preset frequencies
(HOURS -
MINUTES).
PC Connection
:
connect the USB cable inside each package to the USB port on the top panel of the
instrument and the other end to a COM port on the computer.
Use the USB cable supplied with the instrument only.
•
Setup for Data Logger parameter
•
In measure mode press button
to access the SETUP menu.
•
Use the directional keys to move to
LOG SETTINGS P8.0
and access the menu by pressing the key
.
•
Move with the keys
and
to select the program to access.
The table below shows the setup menu structure for the Data Logger mode; for each program, there are
the options that the user can choose and the default value:
•
Composition of setup menu for the Data Logger menu
Program
Description
Options
Factory Default Settings
P8.1
TYPE OF SAVING
MANUAL – HOURS - MINUTES
MANUAL
P8.2
EMPTY INTERNAL MEMORY
YES – NO
-
P8.1 Registration type
Access this menu to select the data acquisition mode:
•
MANUAL
:
The data is acquired only when the user presses the button
.
•
HOURS - MINUTES: Set an automatic data acquisition frequency range.
Use the directional keys to move from MANUAL to HOURS or MINUTES. Access with
and, as