U
SER
M
ANUAL
C
OMMISSIONING
32
Continuous current of the DC power supply
A 12 V DC power supply must deliver a continuous
current of 1 A with 3 A peaks. Use a regulated power
supply or a car battery and note the required voltage and
current levels.
Please note the following sequence:
Connect the power cable to port
(K)
on the Minilogger Z7.
(
Connections and controls)
Tighten the connector screws.
Connect the black plug of the cable to 0 V or ground.
Connect the red plug to the power supply.
The Minilogger Z7 is securely connected to the measuring
environment.
Once the supply voltage is established, the Minilogger Z7 turns on and
displays its operational status via the power button LED
(A)
.
CAUTION
Damage due to improper device shutdown
Disconnecting the power supply during operation may
cause data loss and the destruction of the Minilogger Z7.
Never disconnect the Minilogger Z7 from the power
supply while in operation.
Ensure proper connector seating and tighten the
screws if possible.
Only shut down the Minilogger Z7 through the
X
ORAYA
Suite or the power button on the front panel.
Pull the black plug connected to ground last when
disconnecting the Minilogger Z7 from the power
supply.