OMNIHI5
USER MANUAL
50
MNL-0022
Table 4
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power (W)
Distance
(m)
IMMUNITY
Test Level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM c) ±5kHz
deviation 1
kHz sine
2
0.3
28
710
704-787
LTE Band
13, 17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 85,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700
-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1,3 ,4,
25;UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400
-
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
Modulation
b)
217 Hz
2
0.3
28
5240
5100
-
5800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m
distance is permitted by I EC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.