5.3 Other Functions
5.3.1 Treatment Time Limit
The user can limit the treatment time in the application and the stimulator will stop and
once the time is up. The stimulator will be unavailable until another treatment time limit
is set.
Note: If user doesn
’t want the treatment time to be limited, then the limit value can be set as 0.
5.3.2 Treatment Intensity Limit
The user can set up an intensity limit by using the application for example if the user
sets up the maximum intensity to 25mA, then the maximum electrical output would be
25mA. The limit range is 0-60 mA.
5.3.3 Low Battery Indication
It will flash once per second when the battery is lower than 10%, and the stimulator will
stop and auto off in 5 minutes. The simulator will auto off in 5 seconds when the battery
is lower than 2%.