OMNIHI5
USER MANUAL
47
MNL-0022
Table 2
Guidance and manufacture
’s declaration - electromagnetic emission
The XFT-2003E Nerve and Muscle Stimulator is intended for use in the electromagnetic
environment specified below. The user should comply with specifications ensuring the device
is only used in an appropriate environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±8 kV contact ±2kV,
±4kV, ±8kV,
±15kVair
±8 kV contact ±2kV,
±4kV, ±8kV,
±15kVair
Floors should be
wood, concrete or
ceramic tile.
Humidity should be
at least 30% if it is
synthetic materials.
Electrical fast
transients/bursts
(EFT)
IEC 61000-4-4
±2kV
100kHz repetition
frequency
±2kV
100kHz repetition
frequency
Main power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
±0.5kV,±1kV line-to-
line ±0.5kV,±1kV,
±2kV line-to-ground
±0.5kV,±1kV line-to-
line
±0.5kV,±1kV,±2kV
line-to-ground
Voltage dips
IEC 61000-4-11
0 % UT; 0.5 cycle At
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 0.5 cycle At
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Mains power quality
should be typical
commercial or
hospital environment
UPS power is
recommended if this
device needs to be
used continuously.
0 % UT; 1 cycle and
70 % UT; 25/30
cycles Single phase:
at 0°
0 % UT; 1 cycle and
70 % UT; 25/30
cycles Single phase:
at 0°
Voltage interruptions
IEC 61000-4-11
0% UT; 250/300
cycle
0% UT; 250/300
cycle
RATED power
frequency magnetic
fields IEC 61000-4-8
30A/m
50Hz or60Hz
30A/m
50Hz or 60Hz
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.