OMNIHI5
USER MANUAL
48
MNL-0022
Table 3
Guidance and manufactured declaration - electromagnetic immunity
The XFT-2003E Nerve and Muscle Stimulator is intended for use in the electromagnetic
environment specified below. The user should comply with specifications ensuring the device
is only used in an appropriate environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment
– guidance
Conducted RF
I EC 61000-4-6
3Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF
communications equipment
should be used no closer to any
parts than the recommended
separation distance that
calculated from the equation
applicable to the frequency of
the transmitter.
Recommended separation
distance:
d = 1.2-/P 150 kHz to 80 MHz
d = 1.2VP 80MHzto800 MHz
d = 2.3VP 800MHz to2.7GHz
d = 6VP/E
At RF wireless communications
equipment bands (Portable RF
communications equipment
(including peripherals such as
antenna cables and external
antennas) should be used no
closer than 30 cm (12 inches) to
any part of the device). Where
“P” is the maximum output
power rating of the transmitter in
watts according to transmitter
manufacturer and
“d’ is the
recommended separation
distance in meters. Field
strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
(b), should be less than the
compliance level in each
frequency range (c).Interference
may occur in the vicinity of
equipment marked with the
following symbol:
Radiated RF
I EC 61000-4-3
6Vrmsin ISM
and amateur
radio bands
between 150
kHz and 80 MHz
(a)
10 V/m
80 MHz to 2.7
GHz
6Vrms
10 V/m
Note 1: At 80MHz and 800MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and refection from structures, objects and people.