A
A-8
Color Measurement Instruments
9.
If you have not already done so, remove the shipping spacer from the measurement
page slot.
10.
Insert the end of the Color Reflection Reference strip into the calibration strip
entrance on the DTP41. Center the arrow on the strip below the alignment mark
(see the diagram on
), and slide the strip into the DTP41 until you feel the
resistance of the feed mechanism.
11.
Click Read Reference in the Instrument Calibration dialog box.
The DTP41 pulls the calibration strip through automatically.
A message appears in the Status area when the calibration is finished.
12.
Click Done.
13.
Close the Utilities window.
This completes DTP41 calibration.
When the DTP41 requires calibration, your computer prompts you. When the
prompt message appears, click Calibrate Now and follow the instructions, starting
with
.
After calibrating the DTP41, calibrate the Fiery (see
).
Using the X-Rite DTP32 densitometer
ColorWise Pro Tools are designed to work with the X-Rite DTP32 reflection
densitometer, which feeds color measurements to the Fiery automatically.
N
OTE
:
Measurements from other densitometers can be input using a simple
ASCII file format (see