Guidance and manufacturers declaration – electromagnetic emissions– for all equipment and systems
The STIMPOD NMS 410/450X is intended for use in electromagnetic environment specified below. The customer or user of the STIMPOD NMS 410/450X should assure that it is used
in such an environment
Emission Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
Group 2 – Class A
The STIMPOD NMS 410/450X must emit electromagnetic energy in order to perform its
intended function. Nearby electronic equipment may be affected.
The STIMPOD NMS 410/450X is suitable for use in all establishments, other than
domestic establishments and may be used in domestic establishments and those
directly connected to the public low voltage power supply network that supplies
buildings used for domestic purposes, provided the following warning is heeded:
WARNING: This equipment/system is intended to be used by healthcare professional
only. This equipment/system may cause radio interference or disrupt the operation
of nearby equipment. It may be necessary to take mitigation measures, such as re-
orienting or re-locating the STIMPOD NMS 410/450X or shielding the location
Guidance and manufacturers declaration – electromagnetic immunity- for all equipment and systems
The STIMPOD NMS 410/450X is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS 410/450X should assure that it
is used in such an environment
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 15 kV air
± 6 kV contact
± 15 kV Air
Floors should be wood, concrete or ceramic tile. If the floors are covered with synthetic
material, the relative humidity should be at least 30%
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz
30 A/m (Effective)
Power frequency magnetic fields should be at levels characteristic of a typical location
in a typical commercial or hospital environment.