5.2) Specifications
Operating Modes:
Nerve Locating Mode
Nerve Mapping Mode
NMT Mode
NMS 410/450X
NMS 410/450X
NMS 450X
Current Range
0.00 - 5.00 mA ± 5%
0 - 20mA ± 5%
0 - 80mA ± 5%
Pulse Width Options
0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms ± 5%
0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms ± 5%
0.2ms ± 5%
Maximum Stimulation Voltage
100V
400V
400V
Stimulus
Monophasic square wave
Monophasic square wave
Monophasic square wave
Stimulating Frequency
1Hz , 2Hz, 5Hz ± 5%
1Hz , 2Hz, 5Hz ± 5%
1Hz, 2Hz, 5Hz, 50Hz, 100Hz ± 5%
Load Impedance
0 kOhm - 20 kOhm
0 kOhm - 20 kOhm
0 kOhm - 5 kOhm
Technical Specifications
NMS 410/450X
Device Classification
Class IIa, Type BF
Power Supply
4 x AAA alkaline batteries
Power Consumption
17mA
Waveform
Constant Current, Monophasic Square
Wave
Weight
130g
Dimensions
145mm x 90mm x 30mm
Operating Temperature
10 - 40 ° Celsius
Storage and Transport Temperature
0 - 50 ° Celsius
Operating Humidity
90% Relative Humidity
Transport and Storage Humidity
90% Relative Humidity
Operating Atmospheric Pressure
50 – 106kPa
Transport and Storage Atmospheric Pressure
50 – 106kPa
5.3) Cleaning and Disinfecting STIMPOD NMS 410/450
Cleaning:
Soap and water, applied with a damp cloth is suitable to clean and disinfect the STIMPOD. It is imperative that no moisture penetrates the STIMPOD.
Disinfecting:
Any commercially available methanol - free disinfectant in an ethyl alcohol base can be used for disinfection.