Default Current and Pulse Width Range:
As shown in table 1 in section 4.4 circle the Wheel to select the
predetermined current and pulse width positions sequentially.
NMS 450X
m
en
u
en
ter
NMS 450X
1
2
3
5
6
7
Current Setting
Adjust using wheel
If sound is enabled:
STIMPOD will beep
every time that a stimulus was delivered. The
pitch of the sound will follow the current
intensity level.
Warning:
Notifies the user of a discrepancy
between current setting and average current
of actual stimulus.
Proximity Indicator
Average current of actual stimulus
Stimulating Frequency Setting
Adjust using Enter / Hz width button
Pulse Width Setting
Adjust using Menu / Pulse width button
4
NOTE:
Because the 20 definable positions include both current and pulse width settings,
pulse width cannot be adjusted independently in this mode. This is indicated on the
screen by the fact that pulse width is not highlighted.
2.1b) Adjusting the Current in MAP mode
Current Range: 0 - 20mA adjustable in 1mA increments.
Circle the Wheel to adjust current.
NOTE:
The STIMPOD will automatically default to the Nerve Mapping current range
(0-20mA) and display the ‘ MAP ’ indicator.
When using the Nerve Mapping Probe (NMS 410/450X):