Nerve Locating/Mapping
Mode (NMS 410/450X)
2
1.6) Open Circuit Detection
The Stimpod performs impedance measurements at regular intervals to detect whether
the connection between the STIMPOD and the patient comprises a closed circuit.
Closed Circuit Detected:
•
Stimulation will take place.
•
Stimulating sound will be heard (Single or multiple beeps depending on proximity
indicator setting. Sound pitch will follow the current intensity).
•
The LED stimulus indicator will pulse green with every successful stimulus attempt.
•
The diagnostic screen will provide active feedback on every pulse delivered.
Open Circuit Detected:
•
No stimulation will take place.
•
No stimulating sound will be heard.
•
The LED stimulus indicator will pulse red with every unsuccessful stimulus attempt.
•
A warning screen will appear in the diagnostics screen indicating that an open circuit
was detected.
1.7) Auto Shutdown
STIMPOD will shut down after 10 minutes of no user or patient interaction.
Open Circuit
Detected