Calibration and Troubleshooting
M370046-01
6-15
6
b
Non-isolated Resistive Programming of Current Calibration
To calibrate the non-isolated resistive programming of current:
1. Short the main output with a shunt. Connect a multimeter to measure
the voltage across the shunt.
2. Connect the 4.000 k
Ω
resistor to the non-isolated connector (APG -
J1), one terminal to resistive programming of current (J1.12) and
voltage programming of current (J1.10), and the other terminal to the
common (J1.7).
3. Set the resistive APG level to 4 k
Ω
. The SCPI command is:
SYST:COMM:APR:LEV:CURR 4
4. Set the input source as Resistive Non-Isolated for Current
Programming. The SCPI command is:
SYST:REM:SOUR:CURR ARES
5. Turn the main output to On by pressing the OUTPUT ON/OFF button
on the front panel or send the command:
:OUTP ON
The OUTPUT ON/OFF button will illuminate.
6. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:CURR:RES
7. Press the terminal keys – or + to reduce or increase the main output
current value. Adjust the main output current to 100%.
8. Press any key other than the + or – keys to exit from interactive
calibration mode.
All manuals and user guides at all-guides.com
Summary of Contents for XTR 100-8.5
Page 2: ...All manuals and user guides at all guides com...
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 18: ...xvi All manuals and user guides at all guides com...
Page 30: ...1 8 All manuals and user guides at all guides com...
Page 92: ...3 48 All manuals and user guides at all guides com...
Page 198: ...5 72 All manuals and user guides at all guides com...
Page 262: ...A 36 All manuals and user guides at all guides com...
Page 284: ...IX 4 All manuals and user guides at all guides com...
Page 285: ...All manuals and user guides at all guides com...