Analog Programming Mode
M370046-01
4-13
4
Query for Analog Current Input Level
To query for analog current input level from non-isolated input:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
CAPr
is displayed on the output voltage display.
2. If the mode is currently
nonE
then select any other APG mode. If this
step is done, be sure to restore the APG setting to
nonE
when the
operation is complete. Otherwise, do not rotate the Rotary knob/Enter
button to keep the current APG mode.
3. Press the rotary knob/Enter button to select the APG mode.
The output voltage display should read
LE C
and the output current
display will display APG level.
4. Press the rotary knob/Enter button to return to normal mode.
5. If the APG mode was modified in step 2, execute the procedure to
deactivate the APG voltage mode.
Quick Tip
Remote operation
uses these SCPI
commands. See
page 5–1.
The SCPI command (s) for these instructions are:
[:]SYSTem[<channel>]:REMote:SOURce:CURRent <?|LOCal
|AVOLtage|IAVoltage|ARESistive|IAResistive>
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel:CURRent
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram:CURRent[:DC]?
All manuals and user guides at all-guides.com
Summary of Contents for XTR 100-8.5
Page 2: ...All manuals and user guides at all guides com...
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 18: ...xvi All manuals and user guides at all guides com...
Page 30: ...1 8 All manuals and user guides at all guides com...
Page 92: ...3 48 All manuals and user guides at all guides com...
Page 198: ...5 72 All manuals and user guides at all guides com...
Page 262: ...A 36 All manuals and user guides at all guides com...
Page 284: ...IX 4 All manuals and user guides at all guides com...
Page 285: ...All manuals and user guides at all guides com...