A-18 975-0200-01-01
S
C
PI
C
o
mm
an
d
R
e
fe
re
n
ce
Table A-4
SCPI Commands for Calibration
Function
SCPI Command
Description
Parameter and
Range
Query
Restore
Factory
Calibration
[:]CALibration[<channel>]:RESTore
Restores the calibration to
the factory defaults.
N/A
N/A
Calibrate
OVP
protection.
[:]CALibration[<channel>][:VOLTage]
:PROTection[:OVER]
Calibrate the over voltage
protection.
<OVP Voltage Set
in Volts>
No
Enter Output
Voltage Data
[:]CALibration[<channel>]:OUTPut
:VOLTage
Calibrates the output
voltage.
<Voltage set in
volts>
No
Enter Output
Current Data
[:]CALibration[<channel>]:OUTPut
:CURRent
Set voltage output
calibration level
<Current set in
amps>
No
Calibrate the
Voltage
controlled
Non Isolated
Voltage APG
feature
[:]CALibration[<channel>]:MONitor
[:VOLTage]
Calibrate the voltage
monitor signal system.
? | <NR1> |
<nothing for
interactive mode>
Yes
Calibrate the
Voltage
controlled
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:MONitor
[:VOLTage]:ISOLated
Calibrate the isolated
voltage monitor signal
system.
? | <NR1> |
<nothing for
interactive mode>
Yes
Calibrate the
Voltage
controlled
Non Isolated
Current APG
feature
[:]CALibration[<channel>]:MONitor
:CURRent
Calibrate the current
monitor signal system.
? | <NR1> |
<nothing for
interactive mode>
Yes
Summary of Contents for XTR 100-17
Page 2: ......
Page 3: ...XTR 850 Watt and 1700 Watt Series Programmable DC Power Supply Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 48: ...2 18...
Page 122: ...4 30...
Page 208: ...6 20...
Page 244: ...A 36...
Page 250: ...B 6...
Page 264: ...C 14...
Page 272: ...IX 4...
Page 273: ......