Calibration and Troubleshooting
M370078-01
6-25
6
Storing and Loading Calibration Parameters
Xantrex recommends that you save all the calibration parameters in a text
file so they can be reloaded in the event of a flash failure or a mistake in
calibration.
To get the calibration data, execute the following commands and save
the response in a text file:
The SCPI command (s) to download all the calibration data are:
[:]CALibration:PARameter[:OUTPut]
[:]CALibration:PARameter:APRogram?
[:]CALibration:PARameter:INPut?
The response to these commands are in the 488.2 Arbitrary Program
Block Data format.
If the calibration ever needs to be restored, you can execute the same
commands and write the arbitrary block data response that was received
in the query as the parameter.
For example, downloading the parameter data can be done from the hyper
terminal as follows.
:cal:par?
#262#h2710,#h0000,#h2710,#h0000,#h2710,#h0000,#h2710,
#h0000,#h1AA4
:cal:par:apr?
#259#h7B,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h
7F,#h7F
:cal:par:inp?
#227#h2710,#h00,#h2710,#h00
To restore the XG with a set of previously saved calibration values
can be done as follows.
:cal:par
#262#h2710,#h0000,#h2710,#h0000,#h2710,#h0000,#h2710,#h0
000,#h1AA4
:cal:par:apr
#259#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h
7F,#h7F
:cal:par:inp
#h227#h2710,#h00,#h2710,#h00
Summary of Contents for Watt Series Programmable DC Power Supply XG...
Page 2: ......
Page 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 30: ...1 8...
Page 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Page 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Page 198: ...5 72...
Page 262: ...A 36...
Page 284: ...IX 4...
Page 285: ......