Terminating the
calibration
prematurely
To terminate the calibration process before the scheduled
end, proceed as follows:
Note
If you have not carried out all calibration steps correctly, the
instrument cannot measure with the specified accuracy.
Note
You can switch off the instrument only after you have fin-
ished or terminated the calibration according to the program.
Calibration record
If a printer or PC is connected and switched on, a calibration
record is automatically output (sample printout, see section
4.6.3).
1 Wait until the current calibration step is finished and
the instrument asks you to insert the next calibration
standard.
2 Press the
8
or
2
key until END is displayed in the
lower display line.
3 Press the
1
key.
The instrument returns to the measuring mode (
a
is displayed). If a printer or PC is connected and
switched on, a calibration record is automatically out-
put (see section 4.6.3 C
ALIBRATION
RECORD
).