Maintenance, cleaning, replacement
SenTix
®
FET-D/B
36
l
Insert a new measuring tip according to section 3.3 M
OUNTING
THE
SENSOR
.
5.11
Exchanging the batteries
The batteries are nearly empty if the battery LED flashes. Insert new batteries
as soon as possible. To do so, switch off the sensor and dry it as necessary.
Exchange the batteries according to section 3.2 I
NSERTING
/
EXCHANGING
THE
BATTERIES
).
Warning
To exchange the batteries, first unscrew the cap nut and disconnect the
measuring tip from the converter. Only then remove the converter enclosure.
Thus, you avoid malfunctions.
Always exchange both batteries at the same time.
5.12
Disposal
Probe
We recommend disposing of the sensor without the batteries as electronic
refuse.
Batteries
This note refers to the battery regulation that applies in the Federal Republic
of Germany. We would ask end-consumers in other countries to follow their
local statutory provisions.
Note
The SenTix
®
FET-D/B sensor contains batteries.
Batteries that have been removed must only be disposed of at the recycling
facility set up for this purpose or via the retail outlet.
It is illegal to dispose of them in household refuse.
KCl solution
Disposal according to the safety data sheet.
Cap nut