Troubleshooting
pH 340/ION
61
pH system messages
System message
Cause
Action
0FL
0FL
0FL
0FL
Indication range
exceeded
pH electrode:
•
Not immersed
•
Cable broken
•
Electrolytic gel dried
•
Air bubble in front of
diaphragm
•
Air in diaphragm
pH electrode:
Immerse into sample solution
Exchange electrode
Exchange electrode
Remove air bubble
Suck off air or moisten
diaphragm
E3
E3
E3
E3
Slope or asymmetry
outside of
admissible ranges
(Slope:
-50 ... -62 mV/pH;
Asymmetry:
-30 ... +30 mV)
pH electrode:
•
pH electrode too old
•
Diaphragm contaminated
•
Membrane contaminated
•
Moisture in the plug
•
Electrolytic gel too old
Meter:
•
Wrong calibration procedure
•
Wrong temperature of
solution
(
Without
temperature probe only)
•
Moisture in the socket
Buffer solution:
•
Unsuitable buffer solutions
•
Buffer solutions too old
•
Used up buffer solutions
pH electrode:
Exchange pH electrode
Clean diaphragm
Clean membrane
Dry plug
Exchange electrode
Meter:
Select correct procedure
Set correct temperature
Dry socket
Buffer solution:
Use WTW Technical buffer
solutions
Pay attention to perishability of
solutions.
Use solutions only once.