Operation
pH 3310
34
ba75802d06
05/2017
Sample record
Calibration evaluation
After calibrating, the meter automatically evaluates the calibration.The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.
30.03.2014 15:55
pH 3310
Ser. no. 08502113
CALIBRATION pH
AutoCal TEC
Buffer 1
4.01
Buffer 2
7.00
Buffer 3
10.01
Voltage 1
184.0 mV
Voltage 2
3.0 mV
Voltage 3
-177.0 mV
Temperatur 1
24.0 øC
Temperatur 2
24.0 øC
Temperatur 3
24.0 øC
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...
Display
Calibration
record
Zero point
[mV]
Slope [mV/
pH]
+++
-15 ... +15
-60,5 ... -58
++
-20 ... +20
-58 ... -57
+
-25 ... +25
-61 ... -60.5
or
-57 ... -56
-
-30 ... +30
-62 ... -61
or
-56 ... -50
Clean the electrode according to the
electrode operating manual
Error
Error
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to chapter
6 W