pH 3310
Operation
23
ba75802d06
05/2017
4.3.2
Data storage
This menu contains all functions to display, edit and erase stored mea-
sured values and calibration records.
Note
Detailed information on the memory functions of the pH 3310 is given
in section 4.5.
4.3.3
Automatic
Stability control
The automatic
Stability control
(AutoRead) function continuously
checks the stability of the measurement signal. The stability has a con-
siderable impact on the reproducibility of measured values.
You can activate or switch off the automatic
Stability control
function
(see section 4.4.4).
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when you switch over between the measured parameters with
<M>
.
when the automatic
Stability control
is switched off.
System
/
Interface
/
Output format
ASCII
CSV
Output format for data
transmission
For details, see section 4.6
System
/
Interface
/
Decimal separator
Dot (xx.x)
Comma (xx,x)
Decimal separator
System
/
Interface
/
Output header
Output of a header for
Out-
put format
:
CSV
System
/
Clock
Time
Date
Date format
Settings of time and date.
For details, see section
4.2.4
System
/
Service infor-
mation
Hardware version and soft-
ware version of the meter
are displayed.
System
/
Reset
-
Resets the system settings
to the default values.
For details, see section
4.8.2
Menu item
Setting