Service menu
63
7.3
Displaying the info memory and input signals
7.3.1
Displaying the input signals
Here you find an overview of the current input signals of the different
inputs:
l
Conductivity (a number without dimension is displayed):
(the displayed value refers to the filling level)
e. g. 250 or higher: the filling level electrodes are bypassed
l
analog flow signal (displayed in mA):
e. g. 12: a current of 12 mA flows at the analog input.
l
Event (a number without dimension is displayed):
0 = there is no event signal
1 = there is an event signal
l
digital flow signal (a number without dimension is displayed):
0 = there is no flow signal
1 = there is a flow signal
1
Call up the service menu (see section 7.1 C
2
Using
ud
switch to the INPUT SIGNALS display.
3
Using
ud
switch to the next menu.