Service menu
61
7.2.2
Manually operating the pinching valve
You can manually change the position of the pinching valve.
1
Call up the service menu (see section 7.1 C
2
Using
ud
switch to the PINCH VALVE menu.
3
Press
g
.
The cursor appears in front of the selection.
4
Using
ud
move the cursor to OPEN or CLOSE or OFF.
5
With
g
confirm the setting.
– OPEN/CLOSE: The motor to turn the pinching valve to the
left or right is switched on.
– OFF: The motor to turn the pinching valve is switched off.
The pinching valve remains in the present position.
6
Using
gg
stop the rotation of the pinching valve (corresponds
to confirming the OFF setting).
7
Using
ud
switch to the next selection menu.
or:
Using
e
leave the service menu.