Application range / Delivery equipment
Instruction manual
26
ba31107de03 07/2017
ba31107de05 07/2017
27
Recommended range of application
OxiTop
measuring systems with inductive stirring system, for respirometric
determination of the biochemical oxygen demand (BOD) according to the self-checking
method.
Delivery equipment
Model
OxiTop
IS 6
OxiTop
IS 12
Order Nr.
208 210
208 211
Inductive stirring system
(Stirring device)
6 OxiTop
instruments
12 OxiTop
instruments
Equipment
3 OxiTop
yellow
3 OxiTop
green
6 OxiTop
yellow
3 OxiTop
green
Accessories:
Accessory set 1
(brown sample bottle,
stirring rods, rubber
quivers)
6 x
12 x
Accessory set 2
(stirring rod remover,
overflow measuring
beaker 164 ml and
432 ml, sodium
hydroxide tablets,
note block)
1 x
1 x
Operation of the Single Measuring System
OxiTop
Measuring principle
BOD measurement with the OxiTop
measuring system is based on pressure
measurement (difference measurement). The measuring is made by pressure
measurement via
piezoresistive electronic pressure sensors
.
With the following functions, the OxiTop
measuring system minimizes the measuring
work and is especially suited to the courses of the respirometric BOD measurement:
•
AUTO TEMP function:
Control of the
temperature adjustment
and
automatic start
of the measurement
(1 hour after starting the OxiTop
measuring system at the earliest, after 3 hours at
the latest). It is not necessary to temper the samples to exactly 20°C before starting
OxiTop
. Samples between 15°C and 20°C can be started immediately as BOD
measurement will not begin until a stable 20°C is reached. This is controlled by
the "AutoTemp" feature.
•
Data logging
Automatic measurement with daily storage of measured value for 5 days:
facilitates measurement without supervising, e. g. over the weekend.
•
Current value
Indication of measured value (0 - 40 digits) corresponds to the digits of the
previous WTW BOD instruments. Conversion of mbar to digits is stored in the
instrument, i. e. evaluation of the measured values remains unchanged.
•
Measuring range reserve
from 40 to 50 digits - for overranging without reset by opening the bottle.
Summary of Contents for OxiTop IS 12
Page 1: ...BEDIENUNGSANLEITUNG OPERATING MANUAL ba31107de05 07 2017 OxiTop IS 6 IS 12 ...
Page 2: ...2 ba31107de05 07 2017 ...
Page 14: ...Bedienungsanleitung 14 ba31107de05 07 2017 Inductive Stirring System IS 6 IS 12 ...
Page 19: ...Technische Daten ba31107de05 07 2017 19 Prüfzertifikat EMV IS 6 ...
Page 20: ...Technische Daten 20 ba31107de05 07 2017 Prüfzertifikat EMV IS 12 ...
Page 21: ...Technische Kundenberatung ba31107de05 07 2017 21 ...
Page 22: ...22 ...
Page 34: ...Instruction manual 34 ba31107de03 07 2017 Inductive Stirring System IS 6 IS 12 ...
Page 39: ...Technical data ba31107de03 07 2017 39 Test certificate EMC IS 6 ...
Page 40: ...Technical data 40 ba31107de03 07 2017 Test certificate EMC IS 12 ...
Page 41: ...ba31107de03 07 2017 41 ...
Page 42: ......
Page 43: ......