Cleaning
Cleaning of sample bottles
Do not use disinfectants!
(Disinfectants will kill the required microorganisms!)
Remove
gross contaminations
mechanically, e.g. with a
brush
.
Rinse the bottles with
clear water
or with
water of the next sample
.
(After using detergents rinse thoroughly! Detergent residues may disturb the BOD
5
determination!)
Cleaning of the OxiTop
Single Measuring System
Do not use alcohol or acetone!
Clean with a soft cloth and aqueous soap solution
Battery replacement (approx. every 3 years)
-
Push in spring catch.
-
Open upper lid.
-
Remove batteries.
-
Put in new batteries, pay attention to polarity!
(e.g. WTW-Modell
Batt/Oxitop
, Order No. 209 012).
-
Place upper lid with tap on locking pin (lower lid). Attention! Do not squeeze cable
connection.
-
Close upper lid (make the hook click).
Disposal note:
Please dispose of the lithium batteries at a battery collection point.
Information
Never use sealing grease or other lubricants
for your OxiTop
instruments. Some of these products contain solvents which cause
severe damages on the plastic housings.
The tightness of the BOD bottles is completely sufficient without grease.
Rough contaminations and particles on the contact surfaces of the
rubber quiver annd the OxiTop
must always be wiped off.
WTW grants no guarantee for damages caused by sealing grease.
Labelling
:
To label the bottles clearly, WTW offers a marking set (6 marking rings with
numbers, which can be put on the neck of the bottle):
Model
MARK-6
(6 pcs)
Order No. 209 013.
Check of the measuring system:
On request WTW offer special test instruments for test control.
Further literature:
WTW BOD handbook
DIN 38409T51
DIN 38409T52
Technical data OxiTop
single measuring system
Measuring principle
Respirometric (manometric)
Display
2-digit 7-segment LED, 10 mm
Measured parameter
Digits
Measuring range
0 ... 40 digits (+10 digits for overranging)
Admissible measuring
temperature
20°C
±
1 K
Admissible sample
temperature
(when being filled)
15-20°C
Power supply
Lithium batteries (260 mAh), 2 x type CR 2430 (3V)
Power consumption
Max. 25 mA (during measurement)
Protection class
Protection system
3 IEC 1010
IP 54 IEC 529
EMC Emissions
Immunity
EN 50081-1, FCC Class A
EN 50082-2, NAMUR recommendation
Climate class
Ambient temperature
2, VDI/VDE 3540
Storage:
-25°C ... +65°C
Operation:
+20°C
±
1K
Test mark
CE
Dimensions
Weight
H: 69 mm, diameter 70 mm
Approx. 85 g
Summary of Contents for OxiTop IS 12
Page 1: ...BEDIENUNGSANLEITUNG OPERATING MANUAL ba31107de05 07 2017 OxiTop IS 6 IS 12 ...
Page 2: ...2 ba31107de05 07 2017 ...
Page 14: ...Bedienungsanleitung 14 ba31107de05 07 2017 Inductive Stirring System IS 6 IS 12 ...
Page 19: ...Technische Daten ba31107de05 07 2017 19 Prüfzertifikat EMV IS 6 ...
Page 20: ...Technische Daten 20 ba31107de05 07 2017 Prüfzertifikat EMV IS 12 ...
Page 21: ...Technische Kundenberatung ba31107de05 07 2017 21 ...
Page 22: ...22 ...
Page 34: ...Instruction manual 34 ba31107de03 07 2017 Inductive Stirring System IS 6 IS 12 ...
Page 39: ...Technical data ba31107de03 07 2017 39 Test certificate EMC IS 6 ...
Page 40: ...Technical data 40 ba31107de03 07 2017 Test certificate EMC IS 12 ...
Page 41: ...ba31107de03 07 2017 41 ...
Page 42: ......
Page 43: ......