Calibration and measuring
NitraLyt
®
700 IQ
4 - 24
ba75397d03
12/2004
The calibration procedures,
1 point standard (1)
and
2 point stand. (3)
are carried out in standard solutions. Chloride does not interfere here.
An incorrectly set chloride compensation only affects the following
measurement.
Chloride compensation
for measurements
The function,
Cl compens.
, should always be used if the concentration
of the interfering ion causes the measurement signal to leave the range
of the required accuracy. To avoid calibration and measuring errors, we
recommend to
!
routinely determine the chloride content
!
switch on the
Cl compens.
function
!
enter the chloride content.
The more the actual chloride content deviates from the value entered
for the chloride compensation, the more the displayed measured value
deviates from the current nitrate content.
Fig. 4-7 shows the interrelationship of the real and displayed
NO3-N value for different chloride contents. The characteristic curve
with optimum chloride compensation corresponds to the characteristic
curve without chloride contents.
Fig. 4-7
Influence of chloride on the measured nitrate value
4.2.2
Further influences on the measured value
Greases, oils, certain tensides and similar substances can shorten the
operational lifetime of the NitraLyt
®
NHA/AT exchange electrode.
Therefore, they should not be present in the test sample.
c (NO3-N) [mg/l]
real
c (NO3-N) [mg/l]
angezeigt von NitraLyt 700 IQ
0 mg/l Cl
-
100 mg/l Cl
-
500 mg/l Cl
-
0,1
1
10
100
1000
0,1
10
100
1000
1
Summary of Contents for NitraLyt 700 IQ
Page 1: ...ba75397e03 12 2004 NitraLyt 700 IQ Operating manual IQ SENSOR NET nitrate sensor...
Page 10: ...Safety NitraLyt 700 IQ 2 4 ba75397d03 12 2004...
Page 20: ...Commissioning NitraLyt 700 IQ 3 10 ba75397d03 12 2004...
Page 56: ...Technical data NitraLyt 700 IQ 8 4 ba75397d03 12 2004...
Page 60: ...Indexes NitraLyt 700 IQ 9 4 ba75397d03 12 2004...