General course of a calibration
Preparatory activities
An optimum calibration result is possible if you,
!
before calibrating,
– condition the sensor in the ES/NO3_ISA-50 standard solution
with 50 mg/l NO3-N for approx. 10 min
– set the required calibration procedure (see section 3.4).
!
make sure the temperature of standard solution and test sample is
similar while calibrating
Note
When exchanging an electrode or the entire combination electrode an
initial calibration (see section 4.1.3) always has to be carried out.
Course
Fig. 4-2
Maintenance condition
Note
When the maintenance condition is activated, linked outputs remain in
their current condition. In the measured value display, the measured
value or the condition indicator of the sensor flashes.
Starting the calibration
1
Switch to the measured value display with
m
.
2
Select the measured value display of the required sensor with
d
.
3
Call up the calibration with
c
. The
Maintenance condition:
Linked outputs are frozen.
window appears.
4
Confirm with
g
.
The display appears with a note that an initial calibration may
be necessary.