Commissioning
3 - 5
ba75397d03
12/2004
Caution
Push the electrode into the sensor right up to the stop so that the con-
nection is watertight. Leaks could lead to the destruction of the sensor.
7
Screw the protective hood onto the sensor.
8
Carry out the settings for the sensor on the terminal of the mea-
suring system (see section 3.4).
9
In the case of initial commissioning, condition the sensor
together with the mounted combination electrode in a diluted
standard solution, e.g. ES/NO3_ISA-50 mg/l NO3-N for about
2 hours (see section 6.2).
10
Calibrate the sensor and the electrode with the measuring sys-
tem (see section 4.1).