Calibration and measuring
NitraLyt
®
700 IQ
4 - 16
ba75397d03
12/2004
V
P
:V
S
is the volume ratio between the test sample and standard to be
added. If several combinations for concentration/volume ratio (bars)
are possible for the determination, select the bar where the expected
NO3-N concentration is more in the center.
Entry of volumes
The volume of standard to be added in this calibration procedure has
to be entered on the terminal of the IQ S
ENSOR
N
ET
with an accuracy
of 1/10 ml. Therefore, the volume has to be rounded up or down
accordingly.
Example:
The expected average concentration is approx. 0.9 mg/l NO3-N.
The vertical line in the diagram crosses three ranges:
– Concentration 1 g/l and volume ratio 200:1
– Concentration 1 g/l and volume ratio 100:1
– Concentration 10 g/l and volume ratio 500:1
Selection of the range: Basically, all of the three ranges are suitable.
The lower bar is crossed by the line on the right edge only. There-
fore, the other two ranges are better suited. For further selection,
decide which bar covers the expected measuring range better, i. e.
whether the expected values are rather lower or higher than the
average value. Besides, the decision can be made according to
practical points of view (availability of standard solution and suitable
dosing equipment).
Summary of Contents for NitraLyt 700 IQ
Page 1: ...ba75397e03 12 2004 NitraLyt 700 IQ Operating manual IQ SENSOR NET nitrate sensor...
Page 10: ...Safety NitraLyt 700 IQ 2 4 ba75397d03 12 2004...
Page 20: ...Commissioning NitraLyt 700 IQ 3 10 ba75397d03 12 2004...
Page 56: ...Technical data NitraLyt 700 IQ 8 4 ba75397d03 12 2004...
Page 60: ...Indexes NitraLyt 700 IQ 9 4 ba75397d03 12 2004...