Note
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
Continuing with two-
point calibration
(
Buffer
:
TEC)
5
Wait for the end of the measurement with stability control or
accept the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears.
6
If necessary, finish the calibration procedure as a single-point
calibration with
<M>
.
The calibration record is displayed.
30.09.2009 08:00
30.09.2009 08:00
7
Thoroughly rinse the IDS sensor with distilled water.
8
Immerse the IDS pH sensor in buffer solution 2.
9
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.