Operation
LF 170
110
ba54104e09
11/2010
4.8
Display of instrument information
The following information can be called up:
z
Configuration data. For this, switch to the configuration level. Press
<ESC>
during the password prompt to view all settings in view mode
(see section 4.4.1 C
ALL
UP
CONFIGURATION
LEVEL
).
z
Parameterization data. For this, switch to the parameterization level.
Press
<ESC>
during the password prompt to view all settings in
view mode (see section 4.5.1 C
ALLING
UP
THE
PARAMETERIZATION
LEVEL
).
z
Cell constant
z
Process for temperature compensation
z
Software version
Viewing the cell
constant and software
version
To call up the selected cell constant, press
<ENTER>
in the measured
value view (operating level)
The set process for temperature compensation is also displayed in the
second display line:
– nLF: Temperature compensation using nonlinear temperature
function
– (numeric value) %/K: Temperature compensation using nonlinear
temperature function with the specified temperature coefficients.
– ----: No temperature compensation
Swap to the display of the software version (So) with
<UP><DOWN>
:
Return to the measured value display with
<M>
.
cm
1/
0
n L F
0 0
1
nLF
S o
1
0 0