DIQ/S 181
Technical data
95
ba77083e01
02/2015
Current outputs
(2 x)
Connecting terminals
Cable glands
EMC product and
system characteristics
Output
Galvanically separated from the sensor
Output current
Can be switched between 0 - 20 mA and
4 - 20 mA
In the case of errors, can be set to: 0 ... 21
mA
Max. initial output volt-
age
24 V, in the case of missing or incorrect bur-
den
Accuracy
0.3 % of current value ± 50 µA,
load max. 500
Ω
Functions
Programmable:
– Recorder attenuation, adjustable
0 - 40 mA/s
– Error behavior can be set as required
0 ... 21 mA
– Error behavior according to Namur NE43
can be selected
– Positive and negative characteristic
curve
Terminal type
Screw-type terminal strip, accessible by
raising the lid
Terminal ranges
Solid wires:
Flexible wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
0.2 ... 2.5 mm
2
Suitable for cable diam-
eter
4.5 ... 10 mm or 7 ... 13 mm
EN 61326
EMC requirements for electrical resources
for control technology and laboratory use
– Resources for industrial areas, intended
for indispensable operation
– Interference emission limits for
resources of class B
System lightning pro-
tection
Extended protective characteristics as
opposed to EN 61326
FCC, class A
Summary of Contents for DIQ/S 181
Page 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Page 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Page 103: ......