DIQ/S 181
Outputs
83
ba77083e01
02/2015
Example
Fig. 6-12 Cleaning cycle
Setting
Result
Reference time (h)
:
Reference time
(min)
:
Interval unit
:
Cleaning interval
:
12
0
Hours (h)
8 h
Reference time: 12:00 hours
This specifies the following
start times:
04:00, 12:00 and 20:00 hours
1
Reference time
Start of a cleaning cycle (
t1
)
Start of the specified
Cleaning duration
(
t1a
)
2
End of the specified
Cleaning duration
(
t1a
)
Start of the specified
Adjustment time
(
t1b
)
3
End of the specified
Adjustment time
(
t1b
)
End of the cleaning cycle (
t1
)
4
Reference time ±
Cleaning interval
(
t2
)
Start of a cleaning cycle
t1
Cleaning cycle =
Cleaning duration
(
t1a
)
Adjustment time
(t1b
)
Linked outputs are frozen.
t2
Cleaning interval
relay
condition
open
closed
00:00
Time
3
1
2
t1
t1
t1
t1a t1b
4
4
t2
t2
t2
t2
1
04:00
12:00
24:00
20:00