DIQ/S 181
Installation
21
ba77083e01
02/2015
Mounting the DIQ/S 181
on a top hat rail
Fig. 3-7
Mounting the DIQ/S 181 on a top hat rail
3.6.5
Panel mounting
Materials required
PMS/IQ kit for panel mounting (see chapter 10 A
).
Tools
3 mm set screw wrench (contained in the panel installation kit).
1
Screw the clamping assembly (pos. 1 in Fig. 3-7) on the back
of the universal transmitter with the two plastic tapping screws
(pos. 2).
2
Attach the universal transmitter to the top hat rail from above
using the clamping assembly and press against the rail until
the clamping assembly clicks into place. The universal trans-
mitter can be moved sideways afterwards.
3
To unhook the universal transmitter, press it downward and
pull it forward at the bottom.
Summary of Contents for DIQ/S 181
Page 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Page 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Page 103: ......