Matrix adjustment, check and calibration
4 - 5
ba75651e03
08/2009
4.2.3
Matrix adjustment with the IQ-LabLink
procedure
The IQ-LabLink procedure enables the safe data exchange between
sensors of the online measuring system IQ S
ENSOR
N
ET
and
photometers (e.g. photoLab
®
6000 / spectroFlex - series) with the aid
of a commercial USB memory device.
During the matrix adjustment of the sensor, the measurement data of
the IQ S
ENSOR
N
ET
sensor is synchronized with the photometrically
determined reference data.
Previously, the reference data had to be entered manually during the
matrix adjustment. With the aid of the IQ-LabLink procedure, the
reference data can now be transferred to the IQ S
ENSOR
N
ET
sensors
directly and with no risk of confusion, without manual input by means of
a USB memory device.
System requirement for
the
IQ-LabLink
procedure
z
IQ S
ENSOR
N
ET
:
– Terminal/controller with USB-A interface and software for the IQ-
LabLink procedure (e.g. MIQ/TC 2020 XT)
– Online sensor with software for the IQ-LabLink procedure
(e.g. AmmoLyt
®
Plus
700 IQ)
z
Photometer:
– Photometer with software for the IQ-LabLink procedure
(e.g. photoLab
®
6000 / spectroFlex - series)
Course of the
IQ-LabLink procedure
Note
To carry out the matrix adjustment of an online sensor with the IQ-
LabLink procedure requires operating steps on both instruments:
IQ S
ENSOR
N
ET
system and the photometer (e. g. photoLab
®
6000 /
spectroFlex - series).
The detailed description of the cross instrument operating steps for the
matrix adjustment with the IQ-LabLink procedure on the
IQ S
ENSOR
N
ET
and the photometer is given in an additional operating
manual. This operating manual can be downloaded from the Internet
under www.wtw.com.
Step 1 on the
IQ S
ENSOR
N
ET
terminal
:
Automatic creation of a job file on the USB memory
device with the current sensor values, parameters,
designation of the measuring location and the
automatic allocation of a job reference number for
clear identification
Step 2 on the
photometer:
Automatic recognition of the job files, menu-guided
measurement of all required parameters, storage of
the determined data in the job file
Step 3 on the
IQ S
ENSOR
N
ET
terminal
:
Automatic recognition of the job files, complete
reading of all data required for the matrix adjustment
on keypressing
Summary of Contents for AmmoLyt Plus 700 IQ
Page 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Page 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Page 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Page 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Page 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...