Commissioning
AmmoLyt
®
Plus
700 IQ
3 - 12
ba75651e03
08/2009
Menu item
Selection/values
Explanations
Temp. adjustment
-1.5 °C ... +1.5 °C
The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:
z
Due to the thermal capacity of the
sensor, it is necessary to place it in a
container with at least 2 liters of water.
z
Leave the sensor in this container for at
least 15 minutes while stirring
occasionally, then carry out the
adjustment.
If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.
Save and quit
The system confirms the saving of the
settings and the display switches to the
next higher level.
Quit
The display switches to the next higher
level without saving the new settings.