Matrix adjustment, check and calibration
AmmoLyt
®
Plus
700 IQ
4 - 12
ba75651e03
08/2009
Calibration history
(available in the
IQ S
ENSOR
N
ET
systems
184 XT, 2020 XT and
2020 XT USB only)
In the calibration history, a calibration has the entries
ES1
and
ES2
in
the
Ref1
and
Ref2
columns which the matrix adjustment has not:
Fig. 4-3
Calibration history of
The calibration history contains the following information:
Last electrode
zeroing (here, via
calibration)
Chronological list
with the last matrix
adjustments and
calibrations
Date
Date of the calibration or matrix adjustment
S
Slope [mV] of the electrode.
Note: After a matrix adjustment, the values for the slope are marked
by a star (*) because they were not changed by this procedure.
DV
Drift voltage [mV]
After an electrode zeroing and subsequent calibration or matrix
adjustment, 0 is displayed.
Ref1
/
Ref2
Depending on the procedure.
Matrix adjustment: Reference concentration [mg/l]
Calibration: Used standard solution
K+
Depending on the procedure.
Matrix adjustment: Potassium concentration [mg/l]
Calibration: Standard solution, VARiON
®
/ES-2
P
Procedure (
1
= matrix adjustment,
3
= calibration)
T
Temperature [°C]
R
Evaluation of the calibration or matrix adjustment
+
: Calibration or matrix adjustment.
The sensor measures with the new calibration or adjustment
data.
?
: Calibration or matrix adjustment unsuccessful.
Invalid calibration or adjustment data were discarded.
Measurement is continued with the last valid values.
Summary of Contents for AmmoLyt Plus 700 IQ
Page 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Page 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Page 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Page 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Page 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...