Maintenance and electrode exchange
AmmoLyt
®
Plus
700 IQ
6 - 4
ba75651e03
08/2009
Cleaning the coupling
ring of the protective
hood
The coupling ring can be unscrewed and dismantled for cleaning as
follows:
Fig. 6-2
Dismantling the coupling ring
After the parts have been cleaned, reassemble the coupling ring in
reverse order. When doing so make sure that the tapered side of the
intermediate ring (pos. 2) points towards the sealing ring (pos. 3).
1
Remove the retaining ring (pos. 1 in Fig. 6-2).
2
Remove the intermediate ring (pos. 2) and sealing ring
(pos. 3).
1
2
3