VB1 DM/JAC 2014 v1.0

 

This document is a quick reference guide. For more comprehensive

 

information 

please refer to your Skyla VB1 Operator’s Manual 

 

 

 
Dispensing  the sample 

 

1.  Attach a new pipette tip to the end of the 200ul pipette provided. Do not touch the 

end of the tip. 

2.  Using your thump push down the plunger on the top of the pipette to the stop 

position. 

3.  Immerse the pipette tip below the surface of the sample. 
4.  Slowly release the plunger to draw up the sample, avoid drawing in any air, as this 

will cause the disc to fail. 

5.  Remove the pipette from the sample tube. 
6.  Insert the pipette tip into the sample port. The pipette tip should only make slight 

contact with the bottom of the sample chamber, do not push down hard. 

7.  Keep the reagent disc level and the pipette tip perpendicular to the surface of the 

disc. 

              

 

 

8.  Gently push down the plunger at the top of the pipette to the stop position. All of the 

sample should have been expelled into the sample chamber. Keeping the plunger 
held down gently remove the pipette tip from the sample port. The plunger may now 
be released, remove the disposable tip from the pipette and discard the tip. 

 

 

 

 

9.  Press the 

START 

icon on the touch screen to open the reagent disc 

drawer at the front of the analyzer. 

 

 

 

10. Hold a reagent disc that contains a sample by its edge, with the barcode side 

facing up and the disc kept level. Avoid touching the surface of the reagent 

disc. Gently place the disc into the drawer and press 

OK 

to close the disc 

drawer. 

 

11. Use the touch screen to enter the patient ID (up to 12 numbers) or scan the patient 

ID using the optional external barcode scanner and press 

OK

 

 

 

12. If you press OK without entering a patient ID, the system will use the 

current date and time as the patient ID in the format dd/mm/yyyy hh:mm. 

 

Summary of Contents for Skyla VB1

Page 1: ...1 VB1 DM JAC 2014 v1 0 This document is a quick reference guide For more comprehensive information please refer to your Skyla VB1 Operator s Manual Skyla VB1 Quick Reference Guide...

Page 2: ...rence Guide Contents External features of the analyser 3 VB1 Function Buttons 4 Testing a Sample 5 How to reconstitute a biochemistry control 8 Running a Quality Control 9 Quality Control 9 Interpreti...

Page 3: ...3 VB1 DM JAC 2014 v1 0 This document is a quick reference guide For more comprehensive information please refer to your Skyla VB1 Operator s Manual External features of the analyser...

Page 4: ...cument is a quick reference guide For more comprehensive information please refer to your Skyla VB1 Operator s Manual VB1 Function Buttons Warning To avoid scratching the screen do not use sharp or ha...

Page 5: ...ure Preparing the Disc Discs should be stored at 2 8C and can be used straight from the refrigerator without the need for warming Avoid placing unopened reagent discs in temperatures greater than 25C...

Page 6: ...ular to the surface of the disc 8 Gently push down the plunger at the top of the pipette to the stop position All of the sample should have been expelled into the sample chamber Keeping the plunger he...

Page 7: ...system will store the analysis results and display them on the screen To cancel a test in progress press the CANCEL icon on the screen the system will prompt for confirmation A reagent disc cannot be...

Page 8: ...le holding the serum vial steady on the bench carefully pour all the diluent into the serum vial Do not shake excess diluent out of the diluent vial as this has been accounted for in the dispense 4 Cl...

Page 9: ...L icon The test process is the same as the one for sample analysis however there is no need to input any patient related information 4 The QC control test results are stored separately from sample res...

Page 10: ...Reports Calculated value Test result outside the reference range The test results are significantly impacted by the interference of LIP HEM or ICT of the sample that is outside the permissible range...

Page 11: ...ual Accessing Previous Results The Skyla VB1 automatically stores tests results for display printing or Export to USB To access previous results From the Home Screen press Recall then Search Select th...

Page 12: ...n please refer to your Skyla VB1 Operator s Manual Loading the Printer Paper Open the top cover of the built in thermal printer and place the printer paper into the paper slot Confirm that the heat se...

Page 13: ...etting The system date and time will usually be set before installation and the system remembers the date and time between uses To Change the date and time Select settings from the Home screen Select...

Page 14: ...tomatically cancels the test and displays an operational error message Please insert a new reagent disc to resume operations C The test report exported by the built in thermal printer is blank S The t...

Page 15: ...prevent the drawer from closing properly Contact the authorized technical support department if the issue is not resolved Testing and analysis Internal Memory Full Please contact the authorized techni...

Page 16: ...space is full If the situation is not resolved please contact the authorized technical support department No Firmware Found There is no firmware update file found within the USB drive Please contact t...

Page 17: ...UN CREA Na K B C Liver Panel 7 ALP ALT AST TP ALB TBIL AMY GLOB A G B C Diagnostic Panel 12 ALP ALT AST TB GLU BUN CREA TBIL AMY Ca PHOS GLOB A G B C Skyla VB1 Ordering Information Product Code Descri...

Reviews: