Wolfgang Warmbier GmbH & Co. KG
29
13
Appendix
13.1
Error Messages
The following pop-up windows must be acknowledged or cleared
by pressing any key, and the error must be eliminated.
Pop-Up
Position of the
Function Switch
Function / Meaning
All except for U
Interference voltage
Remedy: device under test must be disconnected from all sources of
voltage
R
ESD
R
OFFSET
> 5
:
OFFSET measurement is not sensible.
Remedy: Check system.
R
ESD
/
Externally accessible fuse is blown.
The voltage ranges remain functional even if the fuse has blown.
Special case for
:
Interference voltage during measurement may result in
a blown fuse.
Remedy: Replace fuse (replacement fuse in battery compartment).
Observe notes regarding fuse replacement in section 12.3.1!
The electronic fuse protects resistance measurements (k
) from over-
loading (electronic hardware circuit).
The voltage ranges remain functional even if the fuse has blown.
All
Battery voltage is less than or equal to 8 V.
Reliable measurement is no longer possible.
Storage of measured values to memory is disabled.
Remedy: Rechargeable batteries must be recharged, or replaced towards
the end of their service life.
Database and Entry Operations
All
The parameters you have selected do not make sense in combination
with previously configured parameters. The selected parameter settings
will not be saved.
Remedy: Enter other parameter settings.
All
Please enter a designation (alphanumeric).
All
Operation with a Barcode Scanner
Error message appears when the “EDIT” entry field is opened and battery
voltage is less than 8 V. Output voltage is generally switched off during
barcode scanner operation if U is less than 8 V in order to assure that
remaining battery capacity is adequate for entering designations for
devices under test and saving the measurement.
Remedy: Rechargeable batteries must be recharged, or replaced towards
the end of their service life.
All
Operation with a Barcode Scanner
Current flowing through the RS 232 port is too high.
Remedy:
The connected device is not suitable for this port.
All
Operation with a Barcode Scanner
Barcode not recognized, incorrect syntax
All
Data cannot be entered at this location within the structure.
Remedy: Observe profile for preselected PC software (see SETUP menu).