background image

18

Wolfgang Warmbier GmbH & Co. KG

6.5

Measurement of Volume Resistance R

V

 according to 

IEC 61340-2-3 

(e.g. for determining the volume resistance of a material)

Connect the red measurement cable from the “+ Pole” to the 
measuring electrode and the black, shielded measurement 
cable with the double “COM / SHIELD” connectors to the 
counter-electrode.

Select a measuring voltage of U

M

 = 10 V in the Resd switch 

position. If a value of less than 1 M

 is displayed, this value is 

the result.

If volume resistance Resd

1 M

 is displayed with the 10 V 

measuring voltage, the 100 V measuring voltage must be se-
lected. The value which is then displayed is the result

Measurement according to IEC 61340-2-3 

6.6

Measurement with the GUARD Terminal

The measurement of very high resistances leads to extremely low 
measuring current and may be rendered problematic as a result 
of influences such as electromagnetic fields, humidity or surface 
current. An accurate test set-up is thus absolutely essential.
A guard cable must be used for measurements within a range of 
10 G

 to 1 T

, in order to prevent surface current from distorting 

measurement results. The GUARD terminal at the measuring 
electrode prevents current on the surface of the high-impedance 
material to be measured from flowing from the + measurement 
cable to the – measurement cable instead of through the high-
impedance material itself.

Ring Electrode according to IEC 61340-2-3

Connect the GUARD terminal at the measuring electrode to 
the GUARD socket at the 

METRISO 3000

 when measuring very 

high-ohmic materials

The measuring sequence is the same as described in the previ-
ous sections.

7

Test Resistor for Resistance Measurement for 
Checking the High Resistance Tester

It’s advisable to check the measuring instrument before, and if 
necessary after use. 
The two outermost jacks on the connection panel must be con-
nected to each other internally via a 10 M

 test resistor to this 

end.
The sum of test resistor and cable resistance (for both cables), 
including test probes, amounts to 10 M

 ±5%. This value allows 

for quick self-testing. 

Connect the measurement cables to the 

+

 and 

COM

 jacks.

Insert the test probes into the above depicted jacks.

Within the R

ESD

 measuring function, use the rotary switch to 

select desired measuring voltage U

M

: 10 V, 100 V or 500 V.

Press the start key and view the measurement results.

Resd in 

Measuring Voltage 

(open-circuit voltage)

Resd 

 1 x 10

6

10 V ± 0.5 V

Resd 

 1 x 10

6

100 V ± 5 V

+Pole

COM/SHIELD

DUT

+Pole

COM/SHIELD

Counter-Electrode

+Pole

GUARD

COM/SHIELD

+

COM

10M

+

COM SHIELD

R

ESD

10 M

Summary of Contents for METRISO 3000

Page 1: ...Operating Instructions METRISO 3000 High Ohmic Resistance Tester for resistance point to point resistance to ground surface and volume resistance measurements 3 349 644 03 2 2 13...

Page 2: ...ED limit value complied with High test voltage with Resd at the MEM Key for memory functions ESC Exit submenu activate instrument from standby state Control Panel Fixed Function Keys LEDs section 4 3...

Page 3: ...aces These connections are located under a protective rubber flap a b Replacement Fuse Socket for Z502R charger Caution Only rechargeable batteries may be inserted when the charger is connected The te...

Page 4: ...Battery OK Battery Level Indicator BAT BAT U 8 5 V Measuring Status Measurement in progress Bars moves from left Measurement on pause Bars are static to right quantities depleted Battery weak Battery...

Page 5: ...ring Probes according to IEC 61340 2 3 17 6 4 2 Measuring Surface Resistance RS of Packaging Materials with a Ring Probe according to IEC 61340 2 3 17 6 5 Measurement of Volume Resistance RV according...

Page 6: ...ents and tests can be performed with the METRISO 3000 Resistance to ground point to point and surface resistance as well as volume resistance with special measuring probes Temperature and atmospheric...

Page 7: ...the measuring circuit mains at all poles Eight 1 5 V size AA batteries in accordance with IEC LR 6 are required for operation of the high impedance measuring instru ment Use new alkaline manganese ba...

Page 8: ...Limit LED test red Cell test Inverse cell test Hide all pixels Show all pixels Acoustic signal test 1 Return to main menu Increase brightness Decrease brightness Increase contrast Decrease contrast S...

Page 9: ...e contrast Decrease contrast Set time Default settings User interface language 3 3a 3b 3c 3d Set date On time for display illumination tester Set time Menu Selection for Operating Parameters Adjusting...

Page 10: ...to a load Note Measuring Sequence If battery voltage drops to below 8 0 V during the course of a measuring sequence this is only indicated by means of a pop up window Measured values are invalid Measu...

Page 11: ...h indicates the detection of interference voltage lights up red the LED is OK In this case the LED reliably indicates interference voltage even when the instrument is switched off We recom mend execut...

Page 12: ...he respective measure ment until is pressed after which the display is returned to the main menu You can return to the main menu by pressing ESC instead of without accepting the newly selected value 1...

Page 13: ...ng point with both test probes Set Voltage Type Parameter waveform Measurement The measured value is displayed directly without pressing the START key in analog format at the bar graph and in digital...

Page 14: ...rs are used only for measurement documentation in the device s internal database and do not influ ence the measurement itself Database functions are explained in sectionl 10 Verification Qualification...

Page 15: ...ce measurement Stop continuous measurement with constant test voltage as soon as the display has settled in Measurement in progress Resd measured resistance U momentary voltage at the test probes Uesd...

Page 16: ...or Rgp System e g for the measurement of a resistance through the combination of a person shoes flooring system according to IEC 61340 4 5 Connect the red measurement cable from the Pole to the measu...

Page 17: ...ng Materials with a Ring Probe according to IEC 61340 2 3 e g for measuring dissipative packaging materials Connect the red measurement cable from the Pole to one measurement jack and the black shield...

Page 18: ...prevents current on the surface of the high impedance material to be measured from flowing from the measurement cable to the measurement cable instead of through the high impedance material itself Rin...

Page 19: ...Temperatures from 10 0 C to 70 0 C and humidity within a range of 10 0 and 90 0 can be measured with the combina tion temperature humidity adapter Observe the operating instructions for the adapter A...

Page 20: ...fer operations are possible Transfer a structure from the PC to the test instrument Transfer a structure including measured values from the test instrument to the PC The test instrument and the PC mus...

Page 21: ...word Search for ID number or text Continue searching Edit menu Cursor LEFT Select an alphanumeric character Cursor RIGHT Select an alphanumeric character ENTER accept an individual character Acknowle...

Page 22: ...accepted and saved in this way are subsequently accepted by the current measuring menu automatically when the display is switched from the struc tural view to measurement Control element parameters c...

Page 23: ...the display to the memory menu Note If you change the parameters in the measurement view they are not saved for the structural element A measurement with changed parameters can nevertheless be saved...

Page 24: ...nners and RFID Readers Search for an Already Scanned Barcode The search can be started from any switch setting and menu Scan the object s barcode The search is started based on the currently selected...

Page 25: ...ally if the measured value remains unchanged for a long period of time and if none of the keys or the rotary switch have been acti vated during on time specified in seconds Service life 3000 measureme...

Page 26: ...ainty Voltage at Device Under Test During Insulation Resistance Measurement Measuring voltage Ux at the device under test depending upon its resistance Rx at nominal voltages of 10 100 and 500 V Figur...

Page 27: ...on during charging Monitor ing of the charging process by the microprocessor might otherwise be disturbed in which case the charging times specified in the technical data can no longer be assured Ple...

Page 28: ...directive In accordance with WEEE 2002 96 EG and ElektroG we identify our electrical and electronic devices as of Aug 2005 with the symbol in accordance with DIN EN 50419 which is shown at the right...

Page 29: ...o memory is disabled Remedy Rechargeable batteries must be recharged or replaced towards the end of their service life Database and Entry Operations All The parameters you have selected do not make se...

Page 30: ...e Remedy Make sure that you have selected the right profile for you PC evaluation program in the SETUP menu see section 3 4 All Memory is full Remedy Save your measurement data to a PC and then clear...

Page 31: ...Wolfgang Warmbier GmbH Co KG 31...

Page 32: ...ithout notice PDF version available on the Internet Wolfgang Warmbier GmbH Co KG Systeme gegen Elektrostatik Untere Gie wiesen 21 78247 Hilzingen Germany Phone 49 77 3186 880 Fax 49 77 3186 8830 e mai...

Reviews: