background image

Wolfgang Warmbier GmbH & Co. KG

7

2

Safety Features and Precautions

The electronic measuring and test instrument is manufactured 
and tested in accordance with safety regulations IEC/EN 61010-
1/VDE 0411-1 and EN 61557. When used for its intended pur-
pose, safety of the operator, as well as that of the instrument, is 
assured.

Read the operating instructions thoroughly and carefully before 
using your instrument. Follow all instructions contained therein.

The measuring and test instrument may not be placed into 
service:

If the battery compartment lid has been removed

If external damage is apparent

If connector cable or measuring adapters are damaged

If  the  instrument  no  longer functions flawlessly

After extraordinary stressing due to transport

After a long period of storage under unfavorable conditions
(e.g. humidity, dust or extreme temperature)

Data Backup

We advise you to regularly transfer your stored data to a PC in 
order to prevent potential loss of data in the test instrument.
We assume no responsibility for any data loss.
We recommend ETC software (Electric Testing Center) for back-
ing up, processing and managing data.

Meanings of Symbols on the Instrument

Warning concerning a point of danger 
(attention, observe documentation!)

Protection class II device

Charging socket for extra-low direct voltage (Z502R 
charger) 

Attention! 
Only rechargeable batteries may be inserted when the char-
ger is connected. 

EC mark of conformity

The device and included batteries may not be dis-
posed of with the trash.

3

Initial Start-Up

3.1

Battery Test

Four different battery symbols, ranging from fully depleted to fully 
charged, continuously indicate the momentary charge level in the 
upper left-hand corner of the display
If battery voltage has fallen below the allowable lower 
limit, the pictograph shown at the right appears. The 
instrument does not function if the batteries have been depleted 
excessively, and no display appears.

3.2

Installing or Replacing Batteries

New batteries must be inserted for initial start-up, or 

if only 1 filled 

segment remains in the battery symbol

.

Attention!

!

Before opening the battery compartment (see page 5 for 
location), disconnect the instrument from the measuring 
circuit (mains) at all poles.

Eight 1.5 V size AA batteries in accordance with IEC LR 6 are 
required for operation of the high-impedance measuring instru-
ment. Use new alkaline manganese batteries only.
Rechargeable NiCd or NiMH batteries may also be used. These 
can be charged externally or by connecting the Z502R charger to 
the test instrument. We recommend rechargeable NiMH batteries. 
Always replace batteries in complete sets.
Dispose of batteries in an environmentally sound fashion.

Loosen both slotted screws for the battery compartment lid 
on the back, and remove the lid.

Remove the battery holder and insert eight 1.5 V size AA bat-
teries with correct polarity in accordance with the symbols. 

Attention!

!

Make sure that 

all of the batteries are inserted with correct 

polarity

. If just one battery is inserted with reversed polar-

ity, it will not be recognized by the instrument and may 
result in leakage from the batteries.

Push the battery holder into the battery compartment such 
that the battery holder’s contacts touch the contact springs at 
the bottom of the battery compartment (see drawing on 
page 3). 
If the battery holder is not inserted as specified, the instrument 
cannot be supplied with power.

Replace the battery compartment lid and retighten the 
screws.

Attention!

!

The instrument may only be placed into service if the 
battery compartment lid is securely fastened!

3.3

Charging the Batteries in the Tester

Attention!

!

Use only the Z502R charger (available as an accessory) 
to charge batteries which have already been inserted into 
the test instrument.

Make sure that the following conditions have been fulfilled be-
fore connecting the charger to the charging socket:

 

– Rechargeable batteries have been installed with 

correct polarity (not standard batteries) 

– The test instrument has been disconnected from the 

measuring circuit at all poles 

– The instrument must remain off during charging.

Refer to section 12.2.1 with regard to charging batteries which 
have been inserted into the tester.

If the batteries or the battery pack have not been used or 
recharged for a lengthy period of time (> 1 month), thus resulting 
in excessive depletion:

Observe the charging sequence (indicated by the LED at the 
charger) and initiate a second charging sequence if necessary 
(disconnect the charger from the mains and from the test instru-
ment to this end, and then reconnect it).

!

Summary of Contents for METRISO 3000

Page 1: ...Operating Instructions METRISO 3000 High Ohmic Resistance Tester for resistance point to point resistance to ground surface and volume resistance measurements 3 349 644 03 2 2 13...

Page 2: ...ED limit value complied with High test voltage with Resd at the MEM Key for memory functions ESC Exit submenu activate instrument from standby state Control Panel Fixed Function Keys LEDs section 4 3...

Page 3: ...aces These connections are located under a protective rubber flap a b Replacement Fuse Socket for Z502R charger Caution Only rechargeable batteries may be inserted when the charger is connected The te...

Page 4: ...Battery OK Battery Level Indicator BAT BAT U 8 5 V Measuring Status Measurement in progress Bars moves from left Measurement on pause Bars are static to right quantities depleted Battery weak Battery...

Page 5: ...ring Probes according to IEC 61340 2 3 17 6 4 2 Measuring Surface Resistance RS of Packaging Materials with a Ring Probe according to IEC 61340 2 3 17 6 5 Measurement of Volume Resistance RV according...

Page 6: ...ents and tests can be performed with the METRISO 3000 Resistance to ground point to point and surface resistance as well as volume resistance with special measuring probes Temperature and atmospheric...

Page 7: ...the measuring circuit mains at all poles Eight 1 5 V size AA batteries in accordance with IEC LR 6 are required for operation of the high impedance measuring instru ment Use new alkaline manganese ba...

Page 8: ...Limit LED test red Cell test Inverse cell test Hide all pixels Show all pixels Acoustic signal test 1 Return to main menu Increase brightness Decrease brightness Increase contrast Decrease contrast S...

Page 9: ...e contrast Decrease contrast Set time Default settings User interface language 3 3a 3b 3c 3d Set date On time for display illumination tester Set time Menu Selection for Operating Parameters Adjusting...

Page 10: ...to a load Note Measuring Sequence If battery voltage drops to below 8 0 V during the course of a measuring sequence this is only indicated by means of a pop up window Measured values are invalid Measu...

Page 11: ...h indicates the detection of interference voltage lights up red the LED is OK In this case the LED reliably indicates interference voltage even when the instrument is switched off We recom mend execut...

Page 12: ...he respective measure ment until is pressed after which the display is returned to the main menu You can return to the main menu by pressing ESC instead of without accepting the newly selected value 1...

Page 13: ...ng point with both test probes Set Voltage Type Parameter waveform Measurement The measured value is displayed directly without pressing the START key in analog format at the bar graph and in digital...

Page 14: ...rs are used only for measurement documentation in the device s internal database and do not influ ence the measurement itself Database functions are explained in sectionl 10 Verification Qualification...

Page 15: ...ce measurement Stop continuous measurement with constant test voltage as soon as the display has settled in Measurement in progress Resd measured resistance U momentary voltage at the test probes Uesd...

Page 16: ...or Rgp System e g for the measurement of a resistance through the combination of a person shoes flooring system according to IEC 61340 4 5 Connect the red measurement cable from the Pole to the measu...

Page 17: ...ng Materials with a Ring Probe according to IEC 61340 2 3 e g for measuring dissipative packaging materials Connect the red measurement cable from the Pole to one measurement jack and the black shield...

Page 18: ...prevents current on the surface of the high impedance material to be measured from flowing from the measurement cable to the measurement cable instead of through the high impedance material itself Rin...

Page 19: ...Temperatures from 10 0 C to 70 0 C and humidity within a range of 10 0 and 90 0 can be measured with the combina tion temperature humidity adapter Observe the operating instructions for the adapter A...

Page 20: ...fer operations are possible Transfer a structure from the PC to the test instrument Transfer a structure including measured values from the test instrument to the PC The test instrument and the PC mus...

Page 21: ...word Search for ID number or text Continue searching Edit menu Cursor LEFT Select an alphanumeric character Cursor RIGHT Select an alphanumeric character ENTER accept an individual character Acknowle...

Page 22: ...accepted and saved in this way are subsequently accepted by the current measuring menu automatically when the display is switched from the struc tural view to measurement Control element parameters c...

Page 23: ...the display to the memory menu Note If you change the parameters in the measurement view they are not saved for the structural element A measurement with changed parameters can nevertheless be saved...

Page 24: ...nners and RFID Readers Search for an Already Scanned Barcode The search can be started from any switch setting and menu Scan the object s barcode The search is started based on the currently selected...

Page 25: ...ally if the measured value remains unchanged for a long period of time and if none of the keys or the rotary switch have been acti vated during on time specified in seconds Service life 3000 measureme...

Page 26: ...ainty Voltage at Device Under Test During Insulation Resistance Measurement Measuring voltage Ux at the device under test depending upon its resistance Rx at nominal voltages of 10 100 and 500 V Figur...

Page 27: ...on during charging Monitor ing of the charging process by the microprocessor might otherwise be disturbed in which case the charging times specified in the technical data can no longer be assured Ple...

Page 28: ...directive In accordance with WEEE 2002 96 EG and ElektroG we identify our electrical and electronic devices as of Aug 2005 with the symbol in accordance with DIN EN 50419 which is shown at the right...

Page 29: ...o memory is disabled Remedy Rechargeable batteries must be recharged or replaced towards the end of their service life Database and Entry Operations All The parameters you have selected do not make se...

Page 30: ...e Remedy Make sure that you have selected the right profile for you PC evaluation program in the SETUP menu see section 3 4 All Memory is full Remedy Save your measurement data to a PC and then clear...

Page 31: ...Wolfgang Warmbier GmbH Co KG 31...

Page 32: ...ithout notice PDF version available on the Internet Wolfgang Warmbier GmbH Co KG Systeme gegen Elektrostatik Untere Gie wiesen 21 78247 Hilzingen Germany Phone 49 77 3186 880 Fax 49 77 3186 8830 e mai...

Reviews: