EMC
Manufacturer's Declaration for Electro-Technical Compatibility
Acc.to IEC 60601-1-2
For the WISAP Cervix Coagulator 6001 incl. Accessories
Guidance and manufacturer's declaration - electromagnetic emissions
The 6001 is intended for use in the electromagnetic environment specified below. The customer or the
user of the 6001 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The 6001 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The 6001 is suitable for use in all establishments
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
RF emissions CISPR 11
Class B
Harmonic
emissions
IEC
61000-3-2
Class A
Voltage
fluctuations/
flicker
emissions IEC 61000-3-3
Not applicable