5
O
PERATION
DANGER!
The WISAP Cervix Coagulator (REF 6001) is a precision instrument, which
is to be operated exclusively with the original WISAP accessories!
WARNING!
The user is fully responsible for observing the applicable cleaning,
disinfection and sterilization regulations. Errors caused by non-observance
of the above regulations are not at the expense of the manufacturer and
exclude any warranty and damage claims.
WARNING!
It is important to ensure that before surgery, cleaning and disinfecting
agents are thoroughly removed.
WARNING!
Do not touch the therapy probe during operation at the distal end!
CAUTION!
Run a visual inspection of the device before each use. When detecting
damage set the unit aside immediately and do not perform any operation
with it. Contact the manufacturer immediately.
NOTE!
Please consult the pertinent medical literature for techniques,
complications and hazards.
5.1 SWITCH ON THE CERVIX COAGUALTOR 6001
Select the correct therapy probe
(4)
required for the intended treatment and join with
receptacle
(2)
of the unit through instrument cable
(3)
.
Rest the Therapy Probe in one of the fixtures for the therapy probes
(11)
and make sure
the tip is not in contact with cloth or any other material.