GA_Portio-Koagulator_E_150914.docx
Page 30 of 44
6
H
YGIENIC MEASURES
In order to maintain effectiveness of the unit, care, maintenance and storage need to be
thoroughly effected. The accessories coming into contact with human tissue need to be
sterilized for avoiding infections to the patient. The different steps described below are
to be performed in the order of their appearance.
6.1 CLEANING
6.1.1 Cleaning of the Unit
After use, the device has to be switched off followed by removing mains cable and all
other accessories. For cleaning the outer surfaces of the unit, an aqueous disinfectant
is suitable. Please observe the manufacturer’s recommendations on solution
concentration. Wipe surface of the unit with a cloth dampened with disinfectant. Entering
of fluid into the unit has to be avoided.
CAUTION!
The controlling unit must not be sterilized!
6.1.2 Cleaning of Therapy Probes and Instrument Cable
Flush therapy probes
4
and instrument cable
3
with running cold and warm water. After
thorough cleaning with demineralized water, the parts have to be dried with sterile
cloths.
6.2 DESINFECTION
6.2.1 Disinfection of Therapy Probes and Instrument Cable
Only parts thoroughly cleaned may be disinfected. Soak parts in solution and observe
the manufacturer’s recommendations on disinfectant concentration and soaking time.
Prolonged soaking time or wrongful concentration may lead to damage at the parts. Do
not pile up parts. Disinfectant residues need to be rinsed off with sterile water under
sterile conditions. Dry all parts with a sterile cloth followed by wrapping them into a sterile
cloth. The disinfected parts need to be stored in a closed, sterile container.