Reference Manual
00809-0100-4160, Rev AB
March 2010
2-23
Rosemount 2160
ADVANCED FUNCTIONS
FOR HART PROTOCOL
Saving, Recalling, and
Cloning Configuration
Data
Use the cloning feature of the Field Communicator or the AMS “User
Configuration” feature to configure several wireless liquid level switches
similarly. Cloning involves configuring a transmitter, saving the configuration
data, then sending a copy of the data to a separate transmitter. Several
possible procedures exist when saving, recalling, and cloning configuration
data. For complete instructions refer to the Field Communicator manual
(publication no. 00809-0100-4276) or AMS on-line guides. One common
method is as follows:
Field Communicator
1. Completely configure the first transmitter.
2. Save the configuration data:
a. Select
F2 SAVE
from the Field Communicator
HOME/ONLINE
screen.
b. Ensure that the location to which the data will be saved is set to
MODULE
. If it is not, select
1: Location
to set the save location to
MODULE
.
c. Select
2: Name
, to name the configuration data. The default is the
transmitter tag number.
d. Ensure that the data type is set to
STANDARD
. If the data type is
NOT
STANDARD
, select
3: Data Type
to set the data type to
STANDARD
.
e. Select
F2 SAVE
.
3. Connect and power the receiving transmitter and Field Communicator.
4. Select the back arrow from the
HOME/ONLINE
screen. The Field
Communicator menu appears.
5. Select
1: Offline
,
2: Saved Configuration
,
1: Module Contents
to
reach the
MODULE CONTENTS
menu.
6. Use the
DOWN ARROW
to scroll through the list of configurations in the
memory module, and use the
RIGHT ARROW
to select and retrieve the
required configuration.
7. Select
1: Edit
.
8. Select
1: Mark All
.
9. Select
F2 SAVE
.
10. Use the
DOWN ARROW
to scroll through the list of configurations in the
memory module, and use the
RIGHT ARROW
to select the configuration
again.
11. Select
3: Send
to download the configuration to the transmitter.
12. Select
OK
after the control loop is set to manual.
13. After the configuration has been sent, select
OK
to acknowledge that the
loop can be returned to automatic control.
When finished, the Field Communicator informs you of the status. Repeat
Steps 3 through 13 to configure another transmitter.
Fast Keys
left arrow, 1, 2
Summary of Contents for Rosemount 2160 WirelessHART
Page 2: ......
Page 18: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 1 10...
Page 56: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 3 14...
Page 70: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 5 10...
Page 76: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 6 6...
Page 92: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 A 16...
Page 98: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 B 6...
Page 99: ......