ASKLÉ
SANTÉ
- 10 -
AXENSOR®
CONNECT:
Details on the operation of the compressor can be transmitted on a PC via a
Bluetooth connection.
Visit the site
www.winncare.fr
, AXENSOR® CONNECT menu to download the software
and find out how it works.
When you stop using the software, check that the compressor starts up again in one of
the three operating modes (DYNAMIC, STATIC or CARE). Otherwise, contact your
maintenance service.
MAINTENANCE
-
DISINFECTION
The method used depends on the degree of disinfection required. Techniques and
products will be applied according to validated good practice recommendations.
See the diagram at the end of the paragraph
- Daily upkeep of the compressor is carried out on the connection between the
mattress and connector.
- It is essential to unplug the electric cable linked to the compressor from the mains.
The light indicating the presence of mains power must be off.
- Do not use high pressure jets to clean an Axtair AUTOMORPHO AXENSOR®
compressor.
- Do not put the compressor on the floor. Hang it at a distance of
≥
60 cm from a
water source.
Do not use corrosive cleaning products such as industrial cleaning agents,
acetone solvents, ether or colouring products (alcohol iodine, potassium
permaganate, silver nitrate, etc.).
Do not use abrasive materials such as steel wool or
Scotchbrite pads.
C
OMPRESSOR
AND
AXENSOR®
SYSTEM
:
Use a dishcloth lightly doused in a CE certified surface detergent or detergent/disinfection solution at
the concentration recommended for use by the manufacturer.
Adhere to the recommended
time periods.
Attention: do not project liquid in order to avoid any damage to the compressor due to
seepage of liquid into the interior of the box.
M
ACRO
-
PARTICLE FILTER
:
The macro-particle filter provided by ASKLE SANTE must be changed at least every 6
months or more frequently, depending on the environmental conditions (dust, smoke,
etc.). This filter is located on the back of the compressor in a transparent cover.