WindSonic Doc No 1405 PS 0019 Issue 25 February 2017
26
Gill Instruments 0.5 metre Aluminium support tube can be supplied with drilled and tapped
M5 holes as above, part number 1405-30-056.
Figure 3 Alignment & Mounting details
WindSonic Doc No 1405 PS 0019 Issue 25 February 2017
27
9 MESSAGE FORMATS
On applying power to the WindSonic, it will provide wind measurements in one of the
following formats:
Gill – Polar, Continuous (
default format
)
Gill - UV, Continuous
Gill – Polar Polled
Gill – UV Polled
NMEA – IIMWV or WIMWV (with NMEA status Codes)
NMEA - IIMWV or WIMWV (with Gill Status Codes)
Tunnel – U Direction, U Speed
Each of these is described below.
Information on how to change the formats and settings follows in Section 10
CONFIGURING.
9.1 Gill format– Polar, Continuous (Default format)
Where:
<STX>
=
Start of string character (ASCII value 2)
WindSonic node address
=
Unit identifier
Wind direction
=
Wind Direction
Wind speed
=
Wind Speed
Units
=
Units of measure (knots, m/s etc.)
Status
=
Anemometer status code (s
ee Section 11.5
for further details)
<ETX>
=
End of string character (ASCII value 3)
Checksum
=
This is the EXCLUSIVE – OR of the bytes
between (and not including) the <STX>
and <ETX> characters.
<CR>
ASCII character
<LF>
ASCII character
<STX>Q, 229, 002.74, M, 00,
<ETX>
16
WindSonic
Node address
Wind
direction
Wind
speed
Units
Status
Check
sum