h. At the Analog Sweep menu, press the menu
soft-key
Sweep Ramp
to go to the Analog
Sweep Ramp menu (shown below).
i. Press
Sweep Time
to open the sweep time pa-
rameter for editing.
j. Set the sweep time to 99 seconds.
NOTE
Monitor the 681XXA’s Horizontal Output on
the Oscilloscope to determine sweep start
and stop
4. During the analog sweep, measure the maximum
and minimum Power Meter readings and record
the values on the test record. Verify that the vari-
ation (difference between the maximum and mini-
mum readings) does not exceed the value noted on
the test record.
OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
5-14
681XXA OM