6-4
ROUTINE MAINTENANCE
Routine maintenance that can be performed by the operator consists of
replacing a defective fuse(s), cleaning the fan filter, and cleaning the
data display.
Replacing a
Fuse
The rear panel line voltage module contains two 5A,
SB, 3AG line fuses. Replace one or both of the line
fuses as follows.
1. On the rear panel, first unplug the power cord
from the power outlet, then unplug it from the
line voltage module.
2. Insert the blade of a small screwdriver into the
slot at the top-center of the line voltage module,
and pry open the cover.
3. Grasp the fuse holders and pull them out (Figure
6-1).
4. Remove the fuses from the fuse holders.
5. Check the fuses and replace any that have failed.
6. Install the fuse holders back into the module,
then close the cover. Ensure the desired voltage
value is displayed through the cover opening.
7. Reinstall the power cord.
OPERATOR
ROUTINE
MAINTENANCE
MAINTENANCE
Fuse Holders
Fuse
Figure 6-1.
Removing/Replacing a Line Fuse
681XXA OM